In-flight calibration of Hitomi Soft X-ray Spectrometer. (3) Effective area
Maeda, Yoshitomo; Tsujimoto, Masahiro; Okajima, Takashi; Eckart, Megan E.; Hayashi, Takayuki; Hoshino, Akio; Iizuka, Ryo; Kelley, Richard L.; Kilbourne, Caroline A.; Leutenegger, Maurice A.; Mori, Hideyuki; Porter, Frederick S.; Sato, Kosuke; Sato, Toshiki; Serlemitsos, Peter J.; Szymkowiak, Andrew; Yaqoob, Tahir
Japan, United States
Abstract
We present the result of the in-flight calibration of the effective area of the Soft X-ray Spectrometer (SXS) on board the Hitomi X-ray satellite using an observation of the Crab nebula. We corrected for artifacts when observing high count rate sources with the X-ray microcalorimeter. We then constructed a spectrum in the 0.5-20 keV band, which we modeled with a single power-law continuum attenuated by interstellar extinction. We evaluated the systematic uncertainty of the spectral parameters by various calibration items. In the 2-12 keV band, the SXS result is consistent with the literature values in flux (2.20 ± 0.08 × 10-8 erg s-1 cm-2 with a 1 σ statistical uncertainty) but is softer in the power-law index (2.19 ± 0.11). The discrepancy is attributable to the systematic uncertainty of about +6%/-7% and +2%/-5% respectively for the flux and the power-law index. The softer spectrum is affected primarily by the systematic uncertainty of the Dewar gate valve transmission and the event screening.